Reading time: about 8 minutes · 19 November 2026 · Science Translational Medicine, 2026

After a spinal cord injury, the hand may still exist, but its touch signals may no longer reach the brain. This paper follows five people who had tiny electrodes placed in the brain’s touch area and asks, over years of use, what happened to their safety, and whether those electrodes could still evoke a sensation of touch.

Background

A spinal cord injury (SCI) can interrupt the pathways that carry movement commands down from the brain and sensory signals back up from the body. Below the level of injury, people may lose voluntary movement and also lose or diminish the sense of touch. For people with tetraplegia — impairment of the arms and legs — restoring more useful hand and arm function is often a primary rehabilitation goal.

An implantable brain–computer interface (iBCI) tries to bypass the damaged pathway. Most motor iBCIs record movement-related activity from the cortex and use it to control a cursor, a robotic arm, or stimulation of muscles. Those systems often still lack the touch feedback that a native hand uses to judge contact and force.

One way to restore a form of touch is intracortical microstimulation (ICMS): brief, very small electrical pulses delivered through microelectrodes implanted in the somatosensory cortex, the cortical region that represents bodily sensation. The part studied here is Brodmann’s area 1, in the hand representation. A pulse train through one electrode can evoke a tactile percept that the person locates on a particular patch of skin. That patch is the electrode’s projected field. When ICMS is combined with a motor iBCI, those percepts can be used as artificial tactile feedback and can improve hand and arm performance.

For such a system to become a clinical device, two questions have to be answered over long periods, not just over weeks. First, what happens to the person: are there serious adverse events — events that are life-threatening, require hospitalization, or cause lasting disability — and are there other, less severe adverse events linked to stimulation? Second, what happens to the sensations and to the electrodes: does the current needed to evoke a feeling rise, do channels stop working, and do the quality and location of the percepts stay usable?

Those questions are easy to mix together. The number of years a device has been in the body is not the same as the number of hours it has actually been stimulating. A count of functional electrodes is not the same as a detection threshold, and neither is the same as whether the evoked sensations still cover the hand or still feel like the same kind of touch. This paper reports those quantities separately.

Problem

Human evidence on the long-term safety and usefulness of ICMS had been limited. Earlier safety observations were much shorter, and reports of human ICMS lasting a few years did not amount to a systematic, multi-person safety and longevity analysis. Animal work had already set conservative stimulation limits, including studies in which earlier, more aggressive protocols could damage tissue near the electrode tip, and later primate work with refined parameters that was used to define the limits for this trial.

The authors therefore looked back across an ongoing early-feasibility trial and asked whether ICMS in human somatosensory cortex, delivered within those limits, remained clinically acceptable and still evoked informative tactile percepts as implants aged — in one person for about a decade, and in others for a few years.

Method

How the stimulation was delivered

Each participant received two Blackrock NeuroPort microelectrode arrays in the hand representation of area 1, and two more arrays in the hand and arm region of motor cortex, all contralateral to the dominant (right) hand (Fig. 1A). The motor arrays were used for recording and as an unstimulated comparison when electrode-health metrics were tracked. ICMS itself was delivered to the sensory arrays with a CereStim R96 stimulator.

Each sensory array is a 6-by-10 grid with 400 μm spacing and 1.5 mm shanks coated in sputtered iridium oxide. Thirty-two electrodes on each array were available for stimulation, so each person had 64 stimulation channels. Pulses were current-controlled, biphasic, and charge-balanced: a 200 μs cathodic phase, a 100 μs interphase, and a 400 μs half-amplitude anodic phase. The maximum frequency and amplitude used in the trial, 300 Hz and 100 μA (20 nC per phase; about 0.5 mC/cm² assuming a 4000 μm² tip), were taken from prior animal studies that motivated the clinical protocol.

A detection threshold is the current at which a person can reliably tell that a pulse train is present. In this paper, an electrode was called functional when its median detection threshold in a given time window stayed below the 100 μA ceiling — that is, when it could still evoke a sensation without exceeding the trial’s amplitude limit. Threshold, functional count, projected-field coverage, and reported quality were analyzed as separate measures.

Figure 1. How to read this figure: panels A–C show the setup. Two arrays sit in area 1 of somatosensory cortex, behind the motor “hand knob.” Pulses at the electrode tips activate nearby neurons and can evoke touch-like percepts on corresponding parts of the hand (the projected fields). Panel D is the implant timeline: P2 in 2015, P3 and C1 around 2020, and P4 and C2 in 2023. Individual follow-up, not the summed 27 years, is what this panel shows. From Greenspon, Hobbs et al., 2026, Figure 1.

Figure 1. How to read this figure: panels A–C show the setup. Two arrays sit in area 1 of somatosensory cortex, behind the motor “hand knob.” Pulses at the electrode tips activate nearby neurons and can evoke touch-like percepts on corresponding parts of the hand (the projected fields). Panel D is the implant timeline: P2 in 2015, P3 and C1 around 2020, and P4 and C2 in 2023. Individual follow-up, not the summed 27 years, is what this panel shows. From Greenspon, Hobbs et al., 2026, Figure 1.

Experiments

The data come from a retrospective analysis of the early-feasibility trial NCT01894802, conducted at the University of Pittsburgh and the University of Chicago under a U.S. FDA Investigational Device Exemption. Safety data were reviewed biannually by an independent data and safety monitoring board. Five men with cervical SCI were implanted between 2015 and 2023. All were older than 25 and right-handed, and all remained implanted at the time of writing.

Their injuries and residual sensation differed. C1 (C4, AIS D) and C2 (C4, AIS D, plus a right brachial plexopathy) had some remaining hand sensation, though C2 was nearly insensate. P2 (C5 motor / C6 sensory, AIS B) and P3 (C6, AIS B) were insensate on the ulnar side of the hand but retained diminished sensation on the thumb and index. P4 (C4, AIS A) had no spared sensation at or below the C5 dermatome.

A central design choice in how the time is counted is to keep individual follow-up separate from cumulative implant duration (person-years). At the time of writing, the five people had been implanted for 33, 34, 63, 65, and 129 months. Those intervals add up to 324 months, or 27 years of combined implant time. They do not mean that any one person was followed for 27 years, and they are not 27 years of continuous stimulation.

Actual stimulation was session-based. Across participants there were 225 to 1390 experimental sessions, and 37% to 74% of those sessions included ICMS, for a combined 623 hours of stimulation, 173 million pulses, and 1724 mC of charge (Table 1 of the paper). Pulse and charge counts also varied widely from person to person:

Participant Months implanted Stimulation hours Pulses (millions) ICMS-related AEs
C1 63 325.9 69.1 3
C2 33 50.1 38.2 3
P2 129 169.4 51.9 25
P3 65 65.7 11.8 16
P4 34 12.2 2.7 6

Safety was scored in several layers. The trial’s primary safety criterion was that the device would not have to be removed for a safety concern in the first year. The paper also reports surgical and implant-hardware events, ICMS-related adverse events, and, in the four participants who retained some natural hand sensation, monofilament thresholds before implantation and at the time of writing. A persistent sensation — a feeling that continued after the pulse train ended — was logged as an adverse event even when it was brief and required no treatment, because it is unintended and could, in a feedback system, be mistaken for contact.

Efficacy and longevity were tracked with regular single-electrode detection thresholds (118 to 2032 measurements per person; first sensations appeared 14 to 111 days after implant), the proportion of functional electrodes, surveys of projected-field location and quality, and electrode-health recordings: signal-to-noise ratio (SNR), peak-to-peak voltage, 1 kHz impedance, and interphase voltage. Those recording metrics were compared between the stimulated sensory arrays and the unstimulated motor arrays, and against the charge delivered to each electrode.

Results

The implantations were described as well tolerated, with no major intraoperative or postoperative complications, and no infection, intracranial hemorrhage, neurological deterioration, or new-onset seizures attributed to the surgery. The first-year explant-for-safety criterion was met in all five people. Two participants later needed a minor revision to resecure a loose percutaneous pedestal; one of those also had minimal skin debridement at the pedestal site. These events were not accompanied by neurological sequelae, systemic infection, or electrode malfunction.

No serious adverse events were associated with ICMS. There were, however, 53 ICMS-related adverse events, all of them persistent sensations (3, 3, 25, 16, and 6 events in C1, C2, P2, P3, and P4). The sensations stayed in the original projected field, were never described as painful, required no medical intervention, and had no lasting effects. Most lasted less than 10 seconds; 36% lasted 10 to 60 seconds; one event in P3 lasted between 2 and 10 minutes, and the longest reported sensation was about 9.5 minutes. On average there was one event per 23,000 stimulation trials, but the rate varied substantially (from about one per 3,000 trials in P4 to one per 79,000 in C1). For three of the five people, events were more common after multielectrode than single-electrode stimulation. Participants with SCI can also have spontaneous residual sensations, and they were asked to report a persistent sensation even when the two were hard to tell apart, so the count may include some false positives.

In the four participants who still had at least partial natural touch in the same parts of the hand, monofilament thresholds did not decrease between the pre-implant test and the time of writing.

Thresholds, functional electrodes, and percepts

Initial detection thresholds were low in four people (median 14.5 to 22.5 μA) and higher in C2 (median 61.5 μA), who also had the longest delay to first sensation. Across electrodes, thresholds rose by an average of 3.67 ± 11.1 μA per year, with significant differences among participants; P4 did not have enough measurements for a slope (Fig. 2, A and B). The abstract rounds this average to about 3.5 μA per year. There was no consistent relationship between the charge delivered to an electrode and its change in threshold.

The functional-electrode count is a different number. One year after implant, more than 75% of electrodes were functional in C1, P2, and P3 (Fig. 2C). At four years, 80% remained functional in C1 and 75% in P3. At 10 years, 60% remained functional in P2, the longest-implanted participant. At the most recent time point, 56% of P4’s electrodes and 48% of C2’s were functional; P4’s thresholds had risen more quickly after the first year, whereas C2’s initially high thresholds later showed improved sensitivity. Across the five people, the most recent proportion of functional electrodes was 64 ± 13%, described in the paper as about a 21% decrease.

Even as some electrodes crossed the 100 μA limit, somatotopic coverage of the hand remained largely unchanged, because projected fields overlapped (Fig. 2D). Quality reports — mostly mechanical (pressure, touch) or paresthetic (tingle, electrical), with some vibration or buzzing — were consistent over time for C1, P2, and P3 (Fig. 2E). Rated naturalness was stable or increased, except in P2, where it began to fall after the eighth year. Two participants regularly reported mild ICMS-associated pain, but those ratings were well below their baseline pain from the injury (median 1.5 versus 7).

Electrode-health recordings

SNR and peak-to-peak voltage tended to decline on both sensory and motor arrays (median SNR slopes from −0.28 to −0.02 per year). The comparison between stimulated and unstimulated arrays was mixed: in one person sensory SNR fell faster, in another motor SNR fell faster, and in the others the difference was not significant. For peak-to-peak voltage, sensory electrodes declined more slowly than motor electrodes in three people. Impedance decayed exponentially on all arrays, without a consistent sensory-versus-motor difference. Interphase voltages drifted in different directions across participants. None of these recording metrics, nor the charge delivered, was consistently related to whether an electrode remained detectable.

Figure 2. How to read this figure: panels A and B are detection thresholds (the current needed to feel a pulse); they can rise even while an electrode is still usable. Panel C is the separate count of functional electrodes (threshold still below 100 μA). The orange trace is P2, still near 60% at 10 years; the most recent values across all five people are the 64 ± 13% figure in the text. Panel D shows that hand coverage can remain similar even as some channels drop out; panel E shows that reported qualities stayed broadly consistent for P2 and C1. Panels F–H are recording and voltage metrics on sensory versus motor arrays, a comparison used to look at electrode health rather than at perception. From Greenspon, Hobbs et al., 2026, Figure 2.

Figure 2. How to read this figure: panels A and B are detection thresholds (the current needed to feel a pulse); they can rise even while an electrode is still usable. Panel C is the separate count of functional electrodes (threshold still below 100 μA). The orange trace is P2, still near 60% at 10 years; the most recent values across all five people are the 64 ± 13% figure in the text. Panel D shows that hand coverage can remain similar even as some channels drop out; panel E shows that reported qualities stayed broadly consistent for P2 and C1. Panels F–H are recording and voltage metrics on sensory versus motor arrays, a comparison used to look at electrode health rather than at perception. From Greenspon, Hobbs et al., 2026, Figure 2.

Analysis

The study’s main contribution is to place several long-term measurements side by side in the same five people, rather than to collapse them into a single safety or success rate.

On safety, the trial observed no serious adverse events linked to ICMS, and no decrease in residual monofilament sensitivity in the four people who still had natural touch in the stimulated skin regions. It also recorded 53 non-serious, ICMS-related persistent sensations, and two later pedestal revisions with one site debridement. Those facts describe the clinical events that were logged under this protocol. Safety here was assessed with event reports, residual sensation tests, and electrode recordings, during laboratory sessions that amounted to 623 hours of stimulation — on the order of tens of minutes per session, several times a week — not with direct inspection of the tissue under the arrays.

On efficacy, a slowly rising average threshold, a falling share of functional electrodes, and relatively stable projected-field coverage and quality can all be true at once. The 64 ± 13% figure is the most recent functional proportion across the five participants. The 60% figure is P2’s proportion after 10 years. C2 and P4, followed for a shorter time, ended near 48% and 56%. The 27 years in the paper are the sum of five implant intervals; the individual stories run from just under three years to a little more than ten.

The recording comparisons suggest that many changes at the electrode–tissue interface tracked implant duration more closely than the amount of ICMS delivered. In that sense, the longevity of the sensations is a property of a chronic implant used at these parameters, not only a property of the pulse count.

Summary

In five people with cervical spinal cord injury, ICMS of the hand area of somatosensory cortex was followed for 33 to 129 months — 27 years of combined implant time — and 173 million pulses. No serious ICMS-related adverse events were recorded; 53 brief persistent sensations were. At the most recent measurements, 64 ± 13% of electrodes still evoked tactile percepts, including 60% after 10 years in one participant, while average detection thresholds rose slowly and the location and quality of the sensations remained largely consistent. Those are separate results, and together they describe what this trial measured over years of laboratory use.

References

  • Greenspon, C. M., Hobbs, T. G., Verbaarschot, C., et al. (2026). Long-term safety and efficacy of intracortical microstimulation in humans. Science Translational Medicine, 18(858), eaec3728. doi.org/10.1126/scitranslmed.aec3728 · PMC
  • Clinical trial: NCT01894802.
  • Figures 1 and 2 are from Greenspon, Hobbs et al. (2026).