A spinal cord injury can leave the circuits that produce walking intact, yet cut the brain’s commands from reaching them. This paper describes a fully implanted wireless bridge that reconnects those two regions in one person, and it keeps two outcomes distinct: what the system can do while it is switched on, and what the person could still do after training when it was switched off.

Background

Walking is not produced only in the brain. When you take a step, executive commands travel down to a network of neurons in the lower, or lumbosacral, spinal cord. Those spinal neurons then activate motor pools — groups of motor neurons — that drive the muscles of the hips, knees and ankles through the sequence needed to accept weight, push off, and swing the leg forward.

Most spinal cord injuries do not destroy those lumbar circuits. They damage the long pathways that carry the brain’s commands down to them. If those descending pathways are interrupted, the walking networks below the injury can still be anatomically present, yet they no longer receive the signals they need. The result is lasting paralysis of the legs.

That distinction matters, because it suggests a possible strategy: if the lumbar circuits are still there, they can be recruited from outside. One way to do this is epidural electrical stimulation. A paddle of electrodes is placed on the outer membrane of the cord (the dura), and brief pulses are delivered to the dorsal root entry zones — the places where sensory nerve roots enter the cord. Stimulating these entry zones can recruit specific leg motor pools. In earlier work, including the STIMO clinical trial, preprogrammed sequences of such stimulation restored standing and basic walking in people with paralysis (Wagner et al., 2018; Rowald et al., 2022).

Those earlier systems did not read the brain. To start each stimulation sequence, they relied on wearable motion sensors that detected leftover movement, or on compensatory strategies such as lifting a heel. Participants could walk in a basic way, but the control did not feel completely natural, and it was hard to adapt the legs to ramps, obstacles, stairs, or a sudden decision to start or stop.

A brain–computer interface takes a different starting point: electrodes record brain activity, and a decoder estimates what the person is trying to do. Here the recording is electrocorticography (ECoG) — an array of electrodes resting on the dura over the cerebral cortex, closer to the neurons than a scalp EEG. The decoded intention is then used to update spinal stimulation in real time. That is a closed-loop motor system: the loop runs from intended movement, through a digital bridge, to stimulation that helps produce the movement. It is not a sensory-feedback prosthesis. The system does not write touch or a sense of limb position back into the brain; any sensation the person has still comes from residual pathways, vision, and the feel of crutches or a walker.

The practical question is whether such a bridge can be made fully implantable, wireless, quick to calibrate, and stable enough for daily use — and whether practising with it can also change what the person can do when the bridge is off.

Problem

The authors set out to restore a direct, volitional link between the brain and the lumbar spinal cord after injury. The aim was not only to trigger a pre-set stepping pattern, but to give the participant continuous control over the timing and amplitude of the muscle activity that makes standing and walking possible, including in settings that change from step to step.

That requires several pieces to work together: a stable recording of cortical activity related to leg movement; a decoder that can be calibrated in minutes and remain reliable; a library of spinal stimulation programs that can target the right motor pools; and a wireless chain that turns decoder output into analogue updates of those programs, fast enough to feel like walking rather than like operating a switch.

Method

The two implants and the wireless bridge

The brain–spine interface (BSI) combines two fully implanted systems with a wearable processing unit (Fig. 1a).

On the recording side, two WIMAGINE devices were placed epidurally over the left and right sensorimotor cortex, above the regions that respond to attempted lower-limb movements. Each implant is a 50 mm titanium case, the thickness of the skull, containing an 8-by-8 grid of 64 electrodes. Once in place, the devices sit flush in the bone and are not visible under the scalp. Because of wireless bandwidth, 32 of the 64 contacts on each implant were used, sampled at 586 Hz. A personalized headset holds two external antennas: one powers the implants by inductive coupling at 13.56 MHz; the other receives the ECoG signals at 402–405 MHz and sends them to a portable base station and processing unit.

On the stimulation side, a 16-electrode Specify 5-6-5 paddle lead sits epidurally over the dorsal root entry zones of the lumbosacral cord. It is driven by an ACTIVA RC implantable pulse generator, a device also used for deep brain stimulation, placed in an abdominal pocket. In this participant, the paddle and pulse generator had already been implanted during the earlier STIMO trial. The pulse generator was upgraded with wireless modules so that stimulation could be adjusted in real time, with a latency of about 100 ms.

The wireless bridge is the processing chain that sits between these two implants. It is not itself a third implant. The processing unit extracts spatial, temporal and spectral features from the ECoG, predicts the intended movement, and converts those predictions into updates of spinal stimulation programs. The updates are sent over a wireless link to the pulse generator.

Figure 1. How to read this figure: panel a is the whole chain, from cortical recording, through a wearable processing unit, to a pulse generator and a 16-electrode spinal paddle that activates leg muscles. Panel b shows how anatomy and magnetoencephalography guided placement of the two 64-channel cortical implants over the leg region of sensorimotor cortex, with postoperative confirmation. Panel c shows the personalized spinal model used to position the paddle over the relevant dorsal root entry zones, again with postoperative images. From Lorach et al., 2023, Figure 1.

Figure 1. How to read this figure: panel a is the whole chain, from cortical recording, through a wearable processing unit, to a pulse generator and a 16-electrode spinal paddle that activates leg muscles. Panel b shows how anatomy and magnetoencephalography guided placement of the two 64-channel cortical implants over the leg region of sensorimotor cortex, with postoperative confirmation. Panel c shows the personalized spinal model used to position the paddle over the relevant dorsal root entry zones, again with postoperative images. From Lorach et al., 2023, Figure 1.

How intention becomes stimulation

The decoder is a recursive, exponentially weighted Markov-switching multilinear model. It makes two kinds of prediction. A gating model, built around a hidden Markov model, estimates the probability that the person intends to move a particular joint — hip, knee or ankle on either side — or to remain at rest. Separate “expert” regression models then estimate the direction and relative amplitude of that intended movement. During calibration, the model can be updated online.

Those predictions do not simply switch stimulation on or off. They proportionally, or analogously, modulate the amplitude of preconfigured stimulation programs, within ranges that had been shown to recruit the intended muscles. For walking, the authors selected programs that target the muscle groups for weight acceptance, propulsion and swing, and linked them to the decoded probabilities. Updated commands were delivered to the pulse generator every 300 ms.

Calibration

Setting up the BSI required two independent procedures (Fig. 2). First, the participant attempted left and right hip, knee and ankle movements while seated, so that the spatial and spectral ECoG features related to each intention could be identified (Fig. 2a). Second, electrode configurations and stimulation parameters — pulse width, frequency and amplitude range — were tuned so that each program recruited the intended muscle group, checked with electromyography (Fig. 2b). The two were then linked, first for a single joint, then for six joints plus rest, and finally for walking.

Figure 2. How to read this figure: panel a maps the cortical features of an attempted left hip flexion; panel b shows how a stimulation configuration is tuned to the left hip flexors, with a polar plot of the resulting muscle responses. In panel c, a short calibration (left) is followed by a test (right): the spectrogram, decoding probability, stimulation amplitude, muscle activity and hip torque rise and fall together, and accuracy climbs to 97% in about 90 seconds. Panels d and e extend the same idea to both legs and three joints, with confusion matrices for decoding and for the muscles actually recruited by stimulation. From Lorach et al., 2023, Figure 2.

Figure 2. How to read this figure: panel a maps the cortical features of an attempted left hip flexion; panel b shows how a stimulation configuration is tuned to the left hip flexors, with a polar plot of the resulting muscle responses. In panel c, a short calibration (left) is followed by a test (right): the spectrogram, decoding probability, stimulation amplitude, muscle activity and hip torque rise and fall together, and accuracy climbs to 97% in about 90 seconds. Panels d and e extend the same idea to both legs and three joints, with confusion matrices for decoding and for the muscles actually recruited by stimulation. From Lorach et al., 2023, Figure 2.

Experiments

The work is a clinical feasibility study, STIMO-BSI (NCT04632290), in a single participant: a 38-year-old man with an incomplete cervical (C5/C6) spinal cord injury from a biking accident ten years earlier. Incomplete here means that some residual pathways through the injury were still present.

He had already completed STIMO (NCT02936453): a five-month rehabilitation programme with targeted epidural stimulation, which allowed him to step with a front-wheel walker when stimulation was on, followed by about three years of home use of stimulation alone. By the time he enrolled in STIMO-BSI, that recovery had plateaued. He had recovered some mobility without stimulation, but he had difficulty going from standing to walking and stopping, could walk only on flat surfaces, and could not adapt his legs to ramps, obstacles or stairs.

The cortical implants were then added. After a two-week rest, there was a six-week mapping and calibration period (30 sessions), then 15 weeks of neurorehabilitation with physiotherapists: 40 sessions of one to three hours. Training emphasized the hip flexors, which were more impaired, and mixed BSI-supported walking, single-joint work, balance, and standard physiotherapy. After this, he could keep using the system at home.

The tests fall into two groups, and they answer different questions.

The first group asks what the BSI does while it is on: calibration accuracy; voluntary hip flexion and walking with the BSI on versus off; walking with the same stimulation programs controlled by foot-mounted motion sensors instead of the brain decoder; walking on ramps, stairs, obstacles and changing terrain; and stability of signals, decoding and stimulation over months, including unsupervised home use with a walker that housed the hardware.

The second group asks what changed after BSI-supported rehabilitation when stimulation was off: lower-limb motor scores, a change in the light-touch sensory score, the Walking Index for Spinal Cord Injury (WISCI II), a six-minute walk test, independent standing time, the timed up-and-go test, the Berg Balance Scale, and walking quality scored from video by physiotherapists who were blinded to the study. The photographs and scores in Figure 4 were all collected without stimulation.

Results

A decoder that could be calibrated in minutes

In the first sessions after surgery, the participant used the decoder to control the hip flexion of an avatar on a screen. With stimulation added, from a lying position, he could control hip-muscle activity and produce torque with 97% accuracy in less than two minutes (97 ± 0.4% after 90 s; Fig. 2c). When the same framework was extended to seven states — left and right hip, knee and ankle, plus rest — gradual control of each joint reached 74 ± 7% accuracy, against a 14% chance level (Fig. 2d,e). The decoder latency for these seven states was 1.1 ± 0.15 s.

For walking, five minutes of calibration were enough for continuous control of hip-flexor activity. During standing, attempts to lift the foot produced about five times more hip-flexor activity with the BSI on than with it off (Fig. 3a).

Walking with the system on

The same configuration supported walking with crutches (Fig. 3b). With the BSI on, decoding of left and right steps rose and fell with intended steps, stimulation amplitudes followed, and the hip flexors and ankle trajectory produced a regular gait. When the BSI was turned off in the middle of walking, the participant immediately lost the ability to take any step, even though the decoder still registered attempts to walk in the cortical signal. Walking resumed as soon as the system was turned back on. He could also choose to start, keep walking, stop, or stand still without false-positive detections that would have disturbed standing. Berg Balance Scale scores were not reduced with the BSI on; they were slightly improved.

Compared with the same stimulation programs driven in a closed loop by motion sensors on the feet, walking on a treadmill with the BSI had gait features closer to those of healthy walkers. The participant also reported a frequent mismatch in timing when stimulation was triggered from the sensors, which made overground walking harder under that condition. With the BSI, he could walk overground independently with crutches.

The same BSI configuration, without retuning for each task, supported walking up and down a steep ramp — about twice as fast as without stimulation — as well as stairs, obstacles, and surfaces that changed underfoot. Outdoors, he could walk with the system and pause voluntarily without spurious step detections (Fig. 3e).

Stability during a year of use, including at home

After a transitional first month in which the spectral content of the ECoG changed modestly, signals remained stable, with an average decrease in spectral power of 0.03 dB per day. The same decoder still supported gradual control of six joints after a two-month gap. Decoding of hip flexions during walking stayed high from the first session after implantation (accuracy 0.92 ± 0.1) through 2, 6 and 11 months after the first activation with updated models (0.93 ± 0.1, 0.97 ± 0.1 and 0.97 ± 0.1; Fig. 3f). Signal quality and walking-decoding accuracy remained globally unchanged over nearly a year. The useful ranges of stimulation amplitude, and the stimulation thresholds, were also stable over a year. The decoder was recalibrated when the participant or the physiotherapists judged it useful, but the decoder features themselves remained stable. Over time, the depth of cortical modulation increased, indicating that the participant became better at modulating his own cortical activity while using the BSI.

For home use, the hardware was built into a walker with a simple tactile interface. Setup, including checking the headset and setting stimulation amplitude bounds, took less than five minutes and could be done without assistance. He used the system regularly for seven months with stable decoding, and rated security, skilfulness and the ability to participate at the maximum of a standard assistive-device questionnaire (PIADS).

Figure 3. How to read this figure: panel a compares a standing hip flexion with the BSI off and on; muscle activity and knee elevation increase when the system is on. Panel b is the key on/off test during walking: stimulation and steps stop when the BSI is switched off, even though cortical attempts are still decoded, and both resume when it is switched back on. Panel e shows outdoor walking and a voluntary pause without false step detections. Panel f tracks decoding of hip flexions from the first sessions through 11 months. From Lorach et al., 2023, Figure 3.

Figure 3. How to read this figure: panel a compares a standing hip flexion with the BSI off and on; muscle activity and knee elevation increase when the system is on. Panel b is the key on/off test during walking: stimulation and steps stop when the BSI is switched off, even though cortical attempts are still decoded, and both resume when it is switched back on. Panel e shows outdoor walking and a voluntary pause without false step detections. Panel f tracks decoding of hip flexions from the first sessions through 11 months. From Lorach et al., 2023, Figure 3.

Recovery measured with the system off

These results describe what the bridge did while it was on. A separate set of measurements, all made without stimulation, asked whether BSI-supported rehabilitation added anything after the earlier plateau (Fig. 4).

After the 40 sessions, voluntary control of the hip flexors, and the hip flexion movements themselves, improved without stimulation (Fig. 4c). This was accompanied by a 4-point gain in the light-touch sensory score and by further change in lower-limb motor scores, which had gone from 23 before STIMO to 29 after STIMO, remained 29 just before BSI rehabilitation, and reached 31 afterwards (Fig. 4d). Standing and walking capacity, scored on WISCI II, rose from 6 before STIMO to 16 after STIMO-BSI. In practical terms, after BSI-supported rehabilitation the participant could walk more than 10 metres with crutches, with no physical assistance and no stimulation. The six-minute walk distance, which had plateaued during years of stimulation-only home use, increased further after this programme (Fig. 4f). Independent standing time increased by 593%, the timed up-and-go improved by 72%, and the Berg Balance Scale improved by 73% (Fig. 4g). Walking quality, scored from the same videos by six blinded physiotherapists, also improved (Fig. 4h).

These off-system changes showed up in daily life: walking independently around the house, getting in and out of a car, or standing with friends at a bar. They are improvements in residual function, measured with crutches or a walker as the tests required. They are not a return to walking without any aid.

Figure 4. How to read this figure: every panel concerns function without stimulation. Panel a compares walking before STIMO, after stimulation-only rehabilitation, and after BSI-supported rehabilitation. Panel b is the timeline of the two trials. Panel c shows larger voluntary hip flexion and flexor muscle activity after the BSI programme. Panels d–h report motor scores, WISCI II, the six-minute walk, standing and balance tests, and blinded observational gait scores. From Lorach et al., 2023, Figure 4.

Figure 4. How to read this figure: every panel concerns function without stimulation. Panel a compares walking before STIMO, after stimulation-only rehabilitation, and after BSI-supported rehabilitation. Panel b is the timeline of the two trials. Panel c shows larger voluntary hip flexion and flexor muscle activity after the BSI programme. Panels d–h report motor scores, WISCI II, the six-minute walk, standing and balance tests, and blinded observational gait scores. From Lorach et al., 2023, Figure 4.

Analysis

The study therefore reports two related but distinct results in the same person.

The first is prosthetic function. While the BSI is on, cortical intention is translated into analogue updates of lumbar epidural stimulation, and that link is sufficient for this participant to stand, walk, climb stairs and handle changing terrain, including outside the laboratory. Turning the system off during early walking tests immediately removed the ability to step, even though the intention was still visible in the cortical recording. At that stage, the digital bridge — not residual descending drive alone — was what made the step happen.

The second is neurological recovery after training. Following a plateau that had lasted through three years of stimulation-only use, a rehabilitation programme that coupled the person’s own intended movements to spinal stimulation produced further gains that were still there when the system was off. The largest motor change was in the hip flexors, which had been the main target of the training. A modest gain in light-touch sensation was also recorded. These residual sensory changes are not evidence that the BSI itself delivered artificial sensory feedback; the implanted loop is a motor bridge.

The word “natural” in the paper is doing specific work. It refers to the participant’s report that he could control the timing and size of his own steps, to the absence of a timing mismatch that had appeared with motion-sensor triggers, and to gait features that were closer to those of uninjured walkers than stimulation alone. It does not mean that walking became equivalent to walking before the injury. After rehabilitation he still used crutches, and many of the clinical walking tests used a walker.

Because residual pathways and the prosthetic stimulation converge on the same lumbar neurons, a continuous match between intention and stimulation is a plausible way to drive further reorganization of the remaining connections between brain and spinal cord. That reading is consistent with the fact that additional recovery appeared only after the BSI programme, once stimulation-only training had already plateaued, and with earlier animal and clinical work on brain-controlled stimulation. The evidence here is from one person with an incomplete cervical injury who already had long experience with epidural stimulation. It shows what was possible in that setting.

Summary

In one person with a chronic incomplete cervical spinal cord injury, a fully implanted, wireless brain–spine interface linked ECoG recordings of intended leg movements to analogue modulation of lumbar epidural stimulation. With the system on, he could stand, walk, climb stairs and move over complex terrain, and he later used the system independently at home for months. After a rehabilitation programme supported by that bridge, he also recovered further residual function: he could walk overground with crutches even when the interface was switched off. The two results should be kept apart. The first is what the digital bridge does while it is running. The second is a partial, but practically meaningful, change in what the body could still do without it.

References

  • Lorach, H., Galvez, A., Spagnolo, V., et al. (2023). Walking naturally after spinal cord injury using a brain–spine interface. Nature, 618, 126–133. doi.org/10.1038/s41586-023-06094-5 · Nature · PMC
  • Wagner, F. B., et al. (2018). Targeted neurotechnology restores walking in humans with spinal cord injury. Nature, 563, 65–71. doi.org/10.1038/s41586-018-0649-2
  • Rowald, A., et al. (2022). Activity-dependent spinal cord neuromodulation rapidly restores trunk and leg motor functions after complete paralysis. Nature Medicine, 28, 260–271. doi.org/10.1038/s41591-021-01663-5
  • Clinical trials: STIMO-BSI, NCT04632290; STIMO, NCT02936453.
  • Figures 1–4 are from Lorach et al. (2023).